Real-Time Three-Dimensional Transesophageal Echocardiography Outperforms Three-Dimensional Printing in Detecting Multiple Perivalvular Leaks after Mitral Valve Replacement:A Case Report and Literature Review
  • +1
  • Zhirong Wang,
  • Qiuxian Wan,
  • Hong You,
  • Qin Wu
Zhirong Wang
The Second Xiangya Hospital of Central South University

Corresponding Author:[email protected]

Author Profile
Qiuxian Wan
The Second Xiangya Hospital of Central South University
Author Profile
Hong You
The Second Xiangya Hospital of Central South University
Author Profile
Qin Wu
The Second Xiangya Hospital of Central South University
Author Profile

Abstract

To report on a patient who developed multiple perivalvular leak (PVL) after a mechanical mitral valve replacement, and how clinicians successfully used real-time three-dimensional transesophageal echocardiography combined with 3D printing to complete transapical perivalvular closure. Additionally, we reviewed the literature to explore the advantages and disadvantages of multimodal imaging for perivalvular